A High-Resolution Small Animal Spect System Developed at Tsinghua

Dai Tiantian,Ma Tianyu,Liu Hui,Cui Junjian,Wei Qingyang,Liu Yaqiang,Wang Shi
DOI: https://doi.org/10.13538/j.1001-8042/nst.22.344-348
2011-01-01
Nuclear Science and Techniques
Abstract:An SPECT system dedicated to small animal imaging shall be of a millimeter spatial resolution or even better. This study was aimed at achieving 0.5-mm spatial resolution for a small animal SPECT system at low cost. It was developed from a single-head clinical SPECT scanner, with a seven-pinhole collimator and a four-degree-of-freedom motion control stage. Several key techniques were developed, including high-resolution image reconstruction algorithm, high accuracy geometrical calibration method, and optimized system matrix derivation scheme. The system matrix was derived from Monte-Carlo simulation and de-noised by fitting each point spread function to a two dimensional Gaussian function. Experiments of point source and ultra micro hot rod phantom were conducted. With a spatial resolution of 0.5-0.6 mm, this system provides a practical way for low-cost high-resolution animal imaging on a clinic SPECT system.
What problem does this paper attempt to address?