A LED-induced capillary fluorescent detection device designed for the space lab

Xiaoqiong Li,Xiaoming Hu,Zhiguang Zhang,Yulin Deng
DOI: https://doi.org/10.1117/12.869788
2010-01-01
Abstract:For the using of space lab, a miniaturized low-power light-emitting diode-induced fluorescence ( LED-IF) detector for capillary electrophoresis was constructed and evaluated. In this work, a more efficient exciting light source based on Led was developed and a convenient detect plate was produced to release the problem of focusing LED light into a spot with a diameter less than 100 mu m. besides these, all closed structure was used to avoid light cluster, and photomultiplier tube was used to be detector. for the reason of power and flexible, the ultra-low Power MCU(MSP430F449) was used to complete signal processing. The system exhibited the result in the concentration range of 1.0x10(-6) to 1x10(-7)Mol/L.