Design and fabrication of a microfluidic detection chip

Bo Yang,Fei Tang,Xiaohao Wang,Zhuo Yang,Jijun Xiong
DOI: https://doi.org/10.19650/j.cnki.cjsi.2009.07.024
2009-01-01
Abstract:A new type of plastic microfluidic detection chip was developed for analyzing and testing samples. The size of the chip is 86 mm×60 mm×4.5 mm. Gas is used to drive the fluid. The chip was fabricated using precision machining method. A simple hydrophilic modification method was applied for microchannels. Experiment results prove that the hydrophilicity is markedly improved. The mechanism of the hydrophilicity improvement is also analyzed. The chip was bonded in room temperature with a dichloroethane-alcohol mixture mixed in a ratio of 1:1. The influence to the bonding result is analyzed for different mixture ratio, bonding time and bonding pressure. Several valves were integrated in the chip. The adhesion technology of the valve membrane was studied.
What problem does this paper attempt to address?