Study on Assistant Laparoscope Surgical Robot to Calibrate Puncture

董九志,张立勋,李艳生
DOI: https://doi.org/10.3969/j.issn.1006-7043.2009.04.015
2009-01-01
Abstract:Based on the requirement of clinical application,an assistant laparoscope surgical robot with active and passive joints was developed.The design of passive posture joints can make the robot accord with the demands for actual operation further.With the character of passive joints,a method to calibrate the relative position of patient's puncture to the basic coordinate system of the robot was provided.Two laser lights of different inclinations were installed below the small arm,and through the geometric relationship the coordinate system of patient's puncture was measured,offering essential laparoscope posture information for the realization of the robot control function.The experiment on calibration system and accuracy analysis showed that patient's puncture can be calibrated precisely by using the method,and the laparoscope surgical robot worked well.
What problem does this paper attempt to address?