Vision-based hand-eye calibration for robot-assisted minimally invasive surgery

Yanwen Sun,Bo Pan,Yongchen Guo,Yili Fu,Guojun Niu
DOI: https://doi.org/10.1007/s11548-020-02245-5
Abstract:Purpose: The knowledge of laparoscope vision can greatly improve the surgical operation room (OR) efficiency. For the vision-based computer-assisted surgery, the hand-eye calibration establishes the coordinate relationship between laparoscope and robot slave arm. While significant advances have been made for hand-eye calibration in recent years, efficient algorithm for minimally invasive surgical robot is still a major challenge. Removing the external calibration object in abdominal environment to estimate the hand-eye transformation is still a critical problem. Methods: We propose a novel hand-eye calibration algorithm to tackle the problem which relies purely on surgical instrument already in the operating scenario for robot-assisted minimally invasive surgery (RMIS). Our model is formed by the geometry information of the surgical instrument and the remote center-of-motion (RCM) constraint. We also enhance the algorithm with stereo laparoscope model. Results: Promising validation of synthetic simulation and experimental surgical robot system have been conducted to evaluate the proposed method. We report results that the proposed method can exhibit the hand-eye calibration without calibration object. Conclusion: Vision-based hand-eye calibration is developed. We demonstrate the feasibility to perform hand-eye calibration by taking advantage of the components of surgical robot system, leading to the efficiency of surgical OR.
What problem does this paper attempt to address?