Abstract:A novel method is presented to measure LED's luminous flux due to its small size and narrow beam. A compound parabolic concentrator is used to collect the lights from a LED in conjunction with a detector which in turn measures the luminous flux. The feasibility is validated by simulating single ray with different angles and 3 types of LEDs with software simulation analysis. Furthermore, 7 LED samples are tested and the performance comparisons with goniophotometer are performed, and the results demonstrate that the method is significantly effective, and all measurement errors are less than 5%.