Comparative of study on detectability of lesions using wide window and conventional window acquisition with dual head coincidence circuit SPECT

He-bei LI,Qian WANG,Yu WANG,Shu-ling QIN,Cai-qun ZHANG,Ya-mei ZHAO
DOI: https://doi.org/10.3321/j.issn:1003-3289.2008.12.040
2008-01-01
Abstract:Objective: To compare lesion detectability of wide peak acquisition and conventional acquisition procedure on dual headed coincidence circuit single photon emission computed tomography (SPECT). Methods: Wide peak acquisition and conventional acquisition procedure on SPECT were used respectively on a phantom study. Visual observation and L/B measurement were used to evaluate the ability of lesion detection of the two procedures. Results: The visual observations of the two procedures were the same. When the contrasts were 2:1 and 2.5:1, the 0.8 cm lesions could not be detected. When the contrasts were not lower than 15:1, spillover appeared. There was no significant difference of the L/B values between the two procedures. Except for the 2.8 cm lesion, the L/B values were all significantly underestimated on the two procedures. The smaller the lesion was, the greater the underestimation of the L/B values was. Conclusion: Wide peak acquisition can shorten the scan time while not affect the lesion detection.