Development of implantable stimulating system used for electroacupuncture

Chuansen Niu,Yue Yu,Hongwei Hao,Luming Li
DOI: https://doi.org/10.1007/978-3-540-79039-6_165
2008-01-01
Abstract:Electroacupuncture (EA) has been proved effective for treatment of various diseases, such as pain, depression and drug abuse in animal experiment and clinical applications. However, professional operations, including acupoints selection and needles insertion, are needed for needle EA and painful sensation may be generated in surface electrode EA because the non-selective stimulation of the flat electrode activates the cutaneous pain receptor. These disadvantages impede their expansion and design of new EA system is motivated. We present an implantable EA system based on a single acupoint, which consists of stimulator implanted into the acupoint and signal transmitter used out of body. The circuit of the internal stimulator is packaged in silicon rubber and only the stimulating electrodes are left out. The internal stimulator is about 1g and its principal size is phi 14mmx4.2mm. The external transmitter is 39mmx43mmxl2mm in size and 20 g in weight. Monomial square pulse is gained and its amplitude is 5mA while the testing load is 500 Omega. The implantable EA system presented in this paper offers a new method of EA stimulation. It can be used for diseases which need long-term, even coercive treatment, such as chronic pain, depression and drug abuse. Because the EA stimulation is applied at a single acupoint and no other acupoints and meridians are activated, so it also provides a feasible way for the study of acupoint specificity.
What problem does this paper attempt to address?