Coded Excitation in Pulsed Wave Doppler Ultrasound Blood Flow Measurements

XIANG Lei,GAO Shangkai
DOI: https://doi.org/10.3321/j.issn:1000-0054.2008.06.030
2008-01-01
Abstract:Coded excitation was used to improve the signal-to-noise ratio and axial resolution for a pulsed wave Doppler ultrasound system for blood flow measurements. Experiments based on a simulation Doppler phantom verified the system performance. Comparison of sonograms using the 7-chip Barker code and the conventional method show that with the same axial resolution, the 7-chip Barker code sonogram penetrated 1 cm further than the conventional method. Also, with the same signal-to-noise ratio, the axial resolution of coded system was much better than the conventional method. Therefore, coded excitation is a better choice than the conventional method when measuring blood flows in small deep vessels and neighboring vessels.