Study on a Digital Doppler Ultrasonic Blood Flow Measurement System

QI Min-Chao,GAO Shang-Kai
DOI: https://doi.org/10.3969/j.issn.0258-8021.2012.03.06
2012-01-01
Abstract:The conventional Doppler ultrasonic blood flow measurement systems use analog circuits to complete the signal acquisition and processing. Such instruments are susceptible to external electromagnetic interference, environment temperature changes and other factors, and it's difficult to implement modern signal processing algorithms to achieve more advanced detections, such as multi-depth detection, power M-mode Doppler and coded excitation. In this paper, a digital Doppler ultrasonic blood flow signal acquisition and processing system was designed, combined with the classic "autocorrelation technology" to achieve multi-depth detection and power M-mode Doppler. And the axial resolution of the system based on the 2-cycle base 13-chip Barker coded excitation was much better than the conventional method. The feasibility of the system was proved by the phantom test. In addition, the human cerebral artery measurement showed that the system could display the blood information on the path of the ultasound wave, and it was useful for the vessel positioning and identification.