Abstract:A detecting system for the transformations of oxyhemoglobin(HbO2) and blood flow of inner ear was presented for the research of motion sickness.Based on the different absorption parameters of HbO2 and deoxyhemoglobin(Hb) in a near infrared spectroscopy(NIRS) at the wavelengths of 660 nm and 810 nm,a set of engineering formulae were derived for detecting HbO2 and blood flow by Lambert-Beer Law.A new sensor was developed to detect and transform two kinds of human physiology parameters:blood oxygen saturation and blood flow of inner ear.By transportating data to a PC,the system could display real-time the relation of two parameters.The experimental results show that the precision of the detecting system is 3% within 20%~100% of blood oxygen saturation,which offers a forceful theoretic proof for the estimation of vertigo.