Design of a Multi-channel Near-Infrared Spectroscopy System Used for Bedside Real-time Monitoring of Newborn Infants' Brain

Feng Yu,Chen Jingsong,Jiao Xuejun,Bai Jing
DOI: https://doi.org/10.16289/j.cnki.1002-0837.2012.01.009
2012-01-01
Abstract:Objective To develop a multi-channel near-infrared spectroscopy system for bedside monitoring blood flow and blood oxygen of newborn infants' brain.Methods Algorithm for the chosen three wavelengths was defined according to modified Lambert-Beer's Law.Integrated circuits were preferred to simplify the system design and make it more portable and stable.With considertion to the feature of newborn infants,a proper source-detector distance was applied and the outline and encapsulation of the probe was optimized.On the premise of safety,forearm cut-off and ValSalva experiments were conducted to validate the system.Results The whole machine weight is less than 0.5 kg and it won't scald the skin when detecting.The probe is flexible enough to fit the curvature of newborn infant's scalp easily.The results of two validating experiments accorded with physiological regularity and they were consistent with previous papers.Conclusion The system is proved to be safe,effective and appropriate for bedside monitoring of newborn infants' brain.
What problem does this paper attempt to address?