Multi-wavelength multi-distance diffuse correlation spectroscopy system for assessment of premature infants’ cerebral hemodynamics

Nikola Otic,John Sunwoo,Yujing Huang,Alyssa Martin,Mitchell B. Robinson,Bernhard Zimmermann,Stefan Carp,Terrie Inder,Mohamed El-Dib,Maria Angela Franceschini,Marco Renna
DOI: https://doi.org/10.1364/boe.505783
IF: 3.4
2024-02-29
Biomedical Optics Express
Abstract:Infants born at an extremely low gestational age (ELGA, < 29 weeks) are at an increased risk of intraventricular hemorrhage (IVH), and there is a need for standalone, safe, easy-to-use tools for monitoring cerebral hemodynamics. We have built a multi-wavelength multi-distance diffuse correlation spectroscopy device (MW-MD-DCS), which utilizes time-multiplexed, long-coherence lasers at 785, 808, and 853 nm, to simultaneously quantify the index of cerebral blood flow (CBF i ) and the hemoglobin oxygen saturation (SO 2 ). We show characterization data on liquid phantoms and demonstrate the system performance on the forearm of healthy adults, as well as clinical data obtained on two preterm infants.
radiology, nuclear medicine & medical imaging,biochemical research methods,optics