X-Ray Spectral Reconstruction Methods Based on Transmission Measurements

Guifen Lin,Limin Luo
DOI: https://doi.org/10.3969/j.issn.1004-9037.2008.01.015
2008-01-01
Abstract:The two X-ray spectral reconstruction methods based on transmission measurements are evaluated. The X-ray spectrum of 10 MeV of maximum energy generated by the medical linear accelerator is simulated as theoretical X-ray spectrum by the Monte Carlo (MO method. The X-ray spectrum is used as the benchmark spectrum. The transmission coefficient T is calculated by the matrix system AF=T. In reverse, the reconstruction spectrum F is determined by solving the matrix system AF=T with the spectra vectorial algebra (SVA) and the biconjugate gradient (BiCG) methods. The reconstruction results show that the BiCG method provides more accurate results than the SVA method compared with the benchmark spectrum. Even the white Gaussian noise is induced into T to simulate measurement errors of the transmission coefficient, the result still exists. The BiCG method provides better results than the SVA method when signal noise ratio (SNR) is higher than 90 dB.
What problem does this paper attempt to address?