Corresponding photon energies in spectral CT and tube voltages in traditional CT using polychromatic X-ray with same CT attenuation values: an in vitro experiment
Ji-hang SUN,Yong LIU,Dong-qian WANG,Hao-yan LI,Fukui RIKA,Tanaka ISAO,Yun PENG
DOI: https://doi.org/10.3969/j.issn.1005-8001.2017.02.004
2017-01-01
Abstract:Objective To find the corresponding photon energies (keV) in dual-energy spectral CT imaging and tube voltages (kVp) in traditional CT using polychromatic X-ray with same CT attenuation values.Methods A 1.2 cm-diameter polypropylene phantom containing 4 inserts with different iodine concentrations (5,10,15,and 20 mg/ml) was used for both dual-energy spectral CT (fast 80 kVp and 140 kVp switching) and traditional CT with polychromatic X-ray (80 kVp,100 kVp,120 kVp,and 140 kVp tube voltages) at the same radiation dose level.With air or soft tissue as background,the attenuation values of the iodine solutions under different kVp in traditional CT and different keV in dual-energy spectral CT were measured using the same regions of interest at the same level.Results The corresponding values were 52±1.0 keV and 80 kVp,58±1.3 keV and 100 kVp,62±1.4 keV and 120 kVp,66±1.3 keV and 140 kVp with air as background;53-±0.8 keV and 80 kVp,59±1.0 keV and 100 kVp,64±1.0 keV and 120 kVp,68± 1.0 keV and 140 kVp with soft tissue background.The corresponding keV did not change significantly with iodine concentration.Conclusion Monochromatic energy level in spectral CT correlates well with kVp in traditional CT using polychromatic X-ray and is independent of iodine concentration.