Convolution Estimation of Spill-Down and Scatter in Dual Isotope (Tc-99M/in-111) Spect

J Feng,BC Penney,XC Pan
DOI: https://doi.org/10.1117/12.465618
2002-01-01
Abstract:Accurate attenuation correction in dual isotope single photon emission computed tomography (SPECT) requires spill-down and scatter correction. Convolution methods that allow such correction were investigated. Three energy windows were used: 1) a Compton scatter window placed below the Tc-99m photopeak window, 2) the Tc-99m photopeak, and 3) both In-111 photopeaks. Spill-down of In-111 into the Tc-99m window was measured using a small source of In-111 at depths of 5, 7, 9, and 11cm. From these data, the convolution kernel relating the In-111 image and the Tc-99m window spill-down image was computed for each depth. Using a small Tc-99m source at the same depths, the convolution kernels describing its spill-down into the Compton window were computed. A factor to scale the Tc-99m spill-down estimate by for obtaining the self-scatter estimate was determined from the total counts in the Tc-99m photopeak vs. the Compton scatter window when primary photons were blocked with a small lead disk. Depths of 5 and 9cm were optimal for spill-down and Tc-99m self-scatter correction, respectively. It was demonstrated that the spill-down and scatter correction increased effective attenuation coefficient towards its "no-scatter" value and dramatically decreased the intensity of an In-111 source in Tc-99m projection images.