MCNP simulations of medical diagnostic X-ray spectra

Yan KONG,Weihai ZHUO,Bo CHEN,Chao ZHAO
DOI: https://doi.org/10.11889/j.0253-3219.2019.hjs.42.110301
2019-01-01
Nuclear Techniques
Abstract:[Background] In medical diagnostic radiology, detailed knowledge of X-ray spectra is required for evaluating the image quality and patients' or staff's radiation dose. [Purpose] This study aims to simulate medical diagnostic X-ray spectra by using Monte Carlo method. [Methods] First of all, a simplified X-ray machine head geometry model was constructed in MCNP (A general Monte Carlo N-Particle transport code) software. Then, the X-ray spectra and surface source data files in the output window were calculated under tube voltage ranging from 50 kVP to 150 kVP, filter ranging from 0.5 mm (Al) to 5 mm (Al) and anode angles ranging from 6°to 14°. [Results] Database of medical diagnostic X-ray spectra and surface sources in the output window are obtained by various machine X-ray head parameters. The calculated spectra are comparable to those of other X-ray spectra software. [Conclusions] X-ray source and its spectra can be correctly simulated by electronically hitting the target through Monte Carlo simulation, providing a basis for future applications.
What problem does this paper attempt to address?