Evaluation of analytic and motion-resistant performance of the Mindray 9006 pulse oximeter.

Liming Peng,Cunliang Yan,Hong Lu,Yong Xia
2007-01-01
Medical Science Monitor
Abstract:Background: The Mindray 9006 is a newly developed pulse oximeter. The performance of noninvasive monitoring ofarterial oxygen saturation by pulse oximetry (SpO(2)) and pulse rate (PR) with the Mindray 9006 was evaluated. Material/Methods: Twenty healthy volunteers participated in this study. The Mindray 9006 was evaluated according to guidelines including HS3-A, EP5-A, and EP9-A2 published by the CLSI, and ISO 9919. A blinded side-by-side comparison of the Mindray 9006 and the Masimo MS-3 oximeter was also performed. Results: The results demonstrated that the coefficients of variation (CV) of impression were less than 1% for SpO(2) and less than 5% for PR analyzed by the Mindray 9006. Comparisons of methods showed there was no significant difference in SIO2/PR analysis between the Mindray 9006 with Rapidlab 855 CO-oxymeter or 812 Electrocardiogram (ECG) monitor (p > 0.05) and between the Mindray 9006 and Masimo MS-3 (p > 0.05). There was good correlation for SpO,, between the Mindray 9006 and CO-oxymetry or the Masimo MS-3 (r > 0.96) and for PR between the Mindray 9006 and ECG or the Masimo MS-3 (r > 0.98). The accuracy as the root-mean-square was less dian 3% for both SpO,, and PR between the Mindray 9006 and CO-oxymetry or between the Mindray 9006 and ECG. Also, there was no significant difference for SpO(2)/PR between the motion and rest conditions (p > 0.05). Conclusions: The overall performance of the Mindray 9006 for SpO(2),, and PR analysis is excellent and compares favorably with the Masimo MS-3.