Large Field Diffusion Weighted Imaging on 3.0T MR: Initial Study

JIANG Xue-xiang
DOI: https://doi.org/10.3321/j.issn:1003-3289.2007.08.040
2007-01-01
Abstract:Objective To evaluate the feasibility and reproducibility of large field diffusion weighted imaging (DWI) on 3.0T MRI. Methods Ten healthy young volunteers were included in this study. In order to evaluate the reproducibility of large field DWI, large field short TI inversion recovery (STIR) DWI were performed twice in healthy volunteers within one week. The scanning extent was 1664 mm from head to leg with b value of 800 s/mm2. The readability of images was evaluated, serious distortion and artifact locations were recorded. Images were postprocessed with Functool 2 software, and apparent diffusion coefficients (ADCs) of partial normal tissues were measured. Paired sample t test was used for statistical analysis. Results Only slight distorsion and ghosting artifacts were found in junction area of neck and thorax on large field DWI. There was good agreement between the mean ADCs measured by two different DWI examinations. Conclusion Large field DWI performed on 3.0T MR scanner is feasible and reproducible.