Energy Re-calibration of HI-13 Tandem Accelerator

LI Zhi-chang,ZHAO Kui,LU Xiu-qin,GUO Ji-yu,CHEN Quan,GUO Gang,LIU Jian-cheng,SHEN Dong-jun,SUI Li,HUI Ning,KONG Fu-quan,CAI Ming-hui,YANG Ming-jian,HU Yue-ming,BAO Yi-wen
DOI: https://doi.org/10.3969/j.issn.1000-6931.2007.06.007
2007-01-01
Abstract:The K constant of magnetic analyzer of HI-13 tandem accelerator is re-calibrated by mains of the resonance nuclear reaction 12C(p,α0)9B in order to avoid the effects of its possible change due to the updating and improving of HI-13.The K value redetermined is(43.528 2±0.000 7) eV·Hz-2,and the energy resolution is ΔE/E=(1.56±0.25)×10-4.