Optimal Design of Random Sparse 2-D Arrays Based on Genetic Algorithm

HUO Jian,YANG Ping,SHI Keren,ZHANG Wei
DOI: https://doi.org/10.3321/j.issn:0371-0025.2006.02.015
2006-01-01
Acta Acustica
Abstract:In order to limit complexity of the 2-D phased array and its 3-D ultrasound imaging system, a optimal design method of random sparse 2-D arrays based on genetic algorithm is developed. Firstly, the aperture size and the element weights of the array reference is pre-determined according to spatial resolution. Secondly, the sparse ratio is determined according to contrast resolution and the sparse array is designed by randomly removing a fraction of the original set of elements from the 2-D array reference. Consequently, the distribution of the reserved element is optimized by the use of the genetic algorithm to achieve acceptable acoustic properties. The acoustic characteristic of the sparse arrays is analyzed through continuous wave theory and pulse ultrasound field model based on the spatial impulse response. Results show that sparse arrays based on genetic algorithm have a good overall performance. The main-lobe is practically unaffected by the random removal of elements, and grating lobes are avoided in these arrays, but the average side-lobe level increases a little. However, it can satisfy the needs for the high-quality ultrasound imaging when the sparse array is optimized using the genetic algorithm.
What problem does this paper attempt to address?