Evaluation of ECG-edit software in CT coronary angiography

WANG Ji-chen,QIU Jian-xing,WANG He,GAO Li,SUN Xiao-wei,HOU Zhen-ya,JIANG Xue-xiang
DOI: https://doi.org/10.3321/j.issn:1003-3289.2006.10.010
2006-01-01
Abstract:Objective: To evaluate the ECG-edit software of Philips Brilliance 64 CT in CT coronary angiography and its influencing factors. Methods: Among the patients with CT coronary angiography in the past 8 months, 28 patients were included in this study as their images were processed with ECG-edit software. The image quality of data bands affected by abnormal ECG was analyzed, and classified into three grades. The paired-sample t test and spearman test of correlation coefficients were used to statistical analysis. Results: There were 22 patients having arrhythmia, and 6 having ECG baseline abnormality. The image quality was affected in 38 data bands of 28 patients. Before processed with ECG-edit software, 36 data bands of image had severe artifact (grade III ), 2 had mild (grade II ). After processed, 28 data bands had no artifact (grade I), 9 had mild, and one had severe artifact. There was significant difference between image quality of before and after ECG- edit processed (P<O. 001). The spearman test revealed that the images quality after ECG-edit software processed was negatively correlated with the patients' heart rate (P=O. 009). Conclusion: ECG-edit software can reduce or obliterate the artifact that was induced by arrhythmia or ECG baseline abnormality, and improve the image quality.
What problem does this paper attempt to address?