Geometrical Parameters Measurement of Contact Lens With A Novel Imaging Method

Weimin Chen,Jie Zhang
2006-01-01
Abstract:An imaging method to measure the geometrical parameters of contact lens is put forward. In imaging method, contact lens is projected into an enlarged image with a projecting lens. An imaging sensor is used to acquire the enlarged image, and a single board computer is adopted to process the image and to do the calculation of parameters. In addition, a glass ring emerged in the liquid was introduced to hold contact lens without any distortion or dehydration. An example of system is provided. With the system, the thickness t can be measured directly, and fore radius r_1 and retral radius r_2 can be calculated with the help of h_1 and h_2 and d_1 and d_2.With an imaging system of 20 times magnification, and a 512 pixels CCD linear imaging sensor of a 50 μm pixel spacing,the system has a resolution of 5 μm, which is much higher than the accuracy requirement of measurement 0.01millimeter.
What problem does this paper attempt to address?