The Installation Of Cosmic-Ray Testing System And The Test Of Resistive Plate Chamber

Fei Lu,Yanlin Ye,Jun Ying,Yong Ban,JiaRong Zheng,Li Xiang Pu,Jianxin Cai,Hongtao Liu,Sijin Qian,Quanjin Wang
DOI: https://doi.org/10.3321/j.issn:0254-3052.2005.07.012
2005-01-01
Abstract:A cosmic-ray testing station (CORTS) in laboratory has been installed, and the performance of a prototype of CMS RE1/2 RPC was tested. The RPC's efficiency is about 85%, the time resolution is 0.87 +/- 0.08ns, and the spatial resolution is 1.5 times the width of read-out strip, which are consistent with the results obtained using the muon beam at CERN. The comparison makes clear that cosmic-ray test can provide reliably the basic parameters which describe the RPC performance.
What problem does this paper attempt to address?