Orthogonal Signal Correction Used For Calibration Transfer Of Fourier Transform Infrared Spectra

Lin Zhang,Liming Zhang,Yan Li,Bingping Liu,LanPing Hu,Junde Wang
DOI: https://doi.org/10.3321/j.issn:0253-3820.2005.12.010
IF: 1.193
2005-01-01
Chinese Journal of Analytical Chemistry
Abstract:Orthogonal signal correction (OSC) was used for the calibration transfer of partial least squares (PLS) model in which four-component gas mixtures were analyzed simultaneously between two fourier trasform infrared (FITR) instruments. Suitable transfer process of OSC, calibration model built with data of one instrument was used directly for the prediction of the data from the second instrument, was determined to obtain good prediction accuracy with the root mean squared error of prediction (RMSEP) of about 10(-3). The transfer results with OSC were compared with those gotten by DS (direct standardization), PDS (piecewise direct standardization), MSC (multiplicative scatter correction) and FIR (finite impulse response).. OSC was characterized of both model simplicity with 4 latent variables of PLS model and robustness with the subset size of 3. RMSEP of DS and PDS were about 10(-3) which was consistent with the one of OSC but it's unnecessary to measure the same samples on different instruments for OSC. The RMSEP of MSC and FIR were about 10(-1) inferior to OSC greatly.