HIGH FREQUENCY ULTRASOUND IMAGE AND TISSUE CHARACTERIZATION FOR HUMAN THYROID

Xiaozhou Liu,Shigong YE,Xiufen Gong,Weiya ZHANG,Yiqing Jia
DOI: https://doi.org/10.3969/j.issn.0258-8021.2001.02.003
2001-01-01
Abstract:Ultrasonic properties of human thyroid were studied from frequency 25MHz to 55MHz with a C-scan ultrasound backscatter microscope system. For quantitative evaluation of tissues, a C-scan microscope system was set up. Three typical acoustic parameters of normal and diseased human thyroid were measured. The results showed that the speed of sound for normal thyroid was 1520m/s, while the sound speed was 1650m/s for the diseased thyroid. At frequency 35MHz, the attenuation coefficient of normal thyroid was 3.4dB/mm, and that of the diseased thyroid was 6.6dB/mm which was almost two times of that of normal thyroid. The backscatter coefficient of the diseased thyroid was 0.00387/(Srmm), which was two and a half times higher than that of normal thyroid (0.00157/(Srmm)). Besides, high resolution ultrasound images were produced with the C-scan system, and the comparison of ultrasound images with histological slice diagrams showed identical results. The measured ultrasound properties were qualitatively related to their histological structure and imaging characterizations.
What problem does this paper attempt to address?