Nasal continuous positive airway pressure treat obstructive sleep apnea syndrome

Shanqun Li,ShanFu Niu,LiYan Jiang,Li Nie,Yin Y. Cai,Chunxue Bai
2001-01-01
Abstract:Purpose: To use nCPAP treat OSAS and evaluat the effect with nCPAP. Methods: We use nCPAP treating 28 patients with OSAS between August 1997 and January 1999. Results: After CPAP treatment, work pressure of CPAP 0.59 - 1.18 kPa(6 - 12)cm H 2O, the state of an illness improve obviously, all kinds of discomfort are disappear. The result of PSG monitor present: the lowerest SaO 2 increased from (56.3 ± 7.5)% to (92.5 ± 8.6)% during sleep, δSaO 2 decreased from (35.6 ± 6.8)% to (6.9 ± 3.7)%; the number of obstructive apnea decreaed from 226.5 ± 147.4 to 10.5 ± 6.1; the number of hypoventilation decreased from 204.5 ± 171.8 to 6.8 ± 4.7, shallow sleep decrease obviously; deep sleep increased obviously; RDI decreased from 62.5 ± 27. 8 to 4.0 ± 2.8. Conclusions: The patients with OSAS have snoring, repeat apnea at night, sleep at day, have a headache and memory go down, inefficiency in work and study, these due to hypoxemia and/or carbon dioxide retention, damage organ function, severity patients have sudden death. CPAP can improve airway obstruct, treatment effect is good. Patient must use right respiratory method, the effective level of CPAP must be determinated by PSG monitor.