Photoacoustic Spectrometer System for Trace Gas Sensing Based Difference-frequency Laser Source

于清旭,Cornelia Fischer,Markus Sigrist
DOI: https://doi.org/10.3321/j.issn:1005-0086.2001.09.014
2001-01-01
Abstract:By mixing an external-cavity diode laser, tunable in 1.5-1.6 μm, and a Q switched Nd:YAG laser at 1.064 μm in periodically poled lithium niobate, a continuously tunable difference frequency generation in 3.35-3.66 μm wavelength region is realized. The difference frequency generator (DFG) has a narrow line width of less then 1 MHz and an average output power of 0.2 mW. Comparing with optical parametric oscillator the DFG has a simple and compact configuration and has no oscillation threshold. Based on the DFG a photoacoustic spectrometer for trace gas sensing has been constructed. With this system photoacoustic spectrums of 0.1% CH4 gas in different pressure has been recorded. The experiment demonstrates the features and the ability of the system for trace gas measurement.
What problem does this paper attempt to address?