Study And Application On Methanol Bacterial Electrode

Xin Li,Xiaohui Wang,Yusheng Sun,Shubo Han
DOI: https://doi.org/10.3321/j.issn:0253-3820.1999.03.007
IF: 1.193
1999-01-01
Chinese Journal of Analytical Chemistry
Abstract:A breed of methanol-utilizing bacteria using methanol as the only carbon source and energy source, obtained from repeated separating and screaning, was immobilized into bacterial membrane(150 mu m) Coupled the immobilized bacterial membrane with a dissolved oxygen electrode, a methanol bacterial electrode sensor was made for the determination of methanol. The sensor was free from the interferences of EtOH, PrOH, BuOH, and so on. It was stable and gave almost constant response throughout repeated use in three months. A linear response within the concentration range of 0.2 similar to 9.3 mg/L(pH = 7.0, 35.0 degrees C, 0.01 mol/L phosphate buffer) and the relative standard deviation of 2.0% in 14 measurements for 1.26 mg/L of methanol were obtained. The response lime by the steady state method was 3 similar to 6 min, and the washing time was about 5 min. The apparent K-m was 55.0 mg/L. The satisfactory results were obtained for the determination of methanol in wine,industrial air and waste water.