Development of measurement for refractive index of liquid by Mach-Zehnder interferometer
Mao-gang HE,Xin-xin HE,Ying ZHANG,Shi ZHANG
DOI: https://doi.org/10.13738/j.issn.1671-8097.2017.02.002
2017-01-01
Journal of Thermal Science and Technology
Abstract:Refractive index of liquid is an important basic datum for physical,chemical,biological,and other disciplines.Interferometry is a useful method for measuring liquid refractive index,using the function between the interference fringe shifts and the rotation angle of the sample.We present a Mach-Zehnder interferometric system to measure liquid refractive index.A route to eliminate the effects of additional interference fringe shifts from liquid cell is presented and a specific data process to check the position of θ=0° is used,which greatly simplifies the process of measurement.We measured the refractive index of distilled water for revising the thickness of liquid sample,and refractive indices of ethanol,cyclohexane and acetone were measured at t=20 ℃ to verify the system.Meanwhile,we analyzed the uncertainty of experiment system,which is in the order of 10-4.