Abstract:A 10mm-diameter miniature annular centrifugal contactor suitable for laboratory use in radiochemical separation has been designed. The tests for both hydraulic performance and mass transfer efficiency have been performed. The 10mm contactor can operate with residence times of 5-20s per stage and holdup volume of 4 mL per stage. It works well for aqueous-organic flow ratio ranging from 0.1 to 15. Europium mass transfer efficiencies of 90% were attained for tributyl phosphate (TBP)-nitric acid system. The laboratory equipment was suitable for remote operation with highly radioactive solutions, and study of fission product behavior in solvent extraction at short residence times. The satisfactory results were obtained in hot tests of TRPO extraction process that were performed by means of the 10 mm miniature centrifugal contactor.