Measurement system of pyroelectric coefficient for pyroelectric material using dynamic current method

Jun Wang,Weizhi Li,Zhiming Wu
DOI: https://doi.org/10.4028/www.scientific.net/AMM.510.232
2014-01-01
Applied Mechanics and Materials
Abstract:A dynamic current method used to measure pyroelectric coefficient (p) for pyroelectric material is described in detail in this paper. The hardware part of the system includes pre-converting circuit, post amplifier circuit, wave-filter circuit, analog-to-digital conversion circuit, temperature test circuit and LCD display panel circuit. The maximum magnification time of amplifier circuit of system is 11300 and the band-stop frequency center for wave-filter circuit is 50 Hz and it's second harmonics 100 Hz. Finally, the measurement system is used to test p parameter of LiTaO3 crystal slice. The lowest p parameter of LiTaO3 crystal with a value of 6.08x10(-9) C/cm(2) degrees C is achieved at 32.5 degrees C.
What problem does this paper attempt to address?