Dosimetric Analysis of Custom-Made Intracavitary Applicator for Nasopharyngeal Carcinoma Treated with High Dose Rate Brachytherapy

李爱菊,祁振宇,伍振宇,曹新平,麦苗青,邓小武
DOI: https://doi.org/10.16073/j.cnki.cjcpt.2013.02.015
2013-01-01
Abstract:OBJECTIVE: To evaluate the dosimetric accuracy of a custom-made intracavitary applicator and explore its use in high dose rate(HDR) brachytherapy of nasopharyngeal carcinoma.METHODS: Ten nasopharyngeal carcinoma patients treated with HDR intracaviatry brachytherapy were selected,each with a custom-made nasopharyngeal applicator.The patients were immobilized on supine position with individual head masks.CT scan was performed during treatment planning and before each treatment fraction.The nasopharyngeal applicator was reconstructed in PLATO brachytherapy planning system(BPS),making it possible to acquire three dimensional coordinates of source dwell positions.The position error of the applicator during the whole treatment sessions was thus assessed by comparing the coordinates of each source position in subsequent CTs with planning CT.The difference of immobilization with head mask vs.without mask was analyzed by using SPSS software.The potential dose error due to position inaccuracy was simulated in PLATO BPS.RESULTS:For patients without head immobilization mask,the mean position error of the distal end of the applicator was(1.62±0.24),(2.32±0.22) and(2.50±0.28)mm for P1 point,(1.13±0.11),(1.22±0.20) and(1.80±0.15) mm for P5 point,in X,Y,and Z direction.For patients immobilized with head mask,the corresponding displacement of the applicator was on average of(0.82±0.06),(0.70±0.07) and(0.94±0.08) mm for P1 point,(0.70±0.02),(0.46±0.01) and(0.60±0.03) mm for P5 point.Statistically significant difference was found between these two immobilization methods(P0.01).The potential dose error due to position inaccuracy was within 1.2% at dose points of interest(2 cm away from the source) for patients with immobilization mask.CONCLUSIONS:The application of the custom-made nasopharyngeal applicator with the immobilization mask can significantly improve the dosimetric accuracy of HDR intracavitary brachytherapy.The method described here is universal and can be conveniently implemented in clinic.
What problem does this paper attempt to address?