Calibration method for automatic optical measurement instrument of precipitation based on linear CCD

Taichang GAO,Xiaoyong SU,Shuchen YANG,Xichuan LIU
DOI: https://doi.org/10.3969/j.issn.1009-3443.2012.03.190
2013-01-01
Abstract:To prompt the accuracy of automatic optical measurement instrument of precipitation, the emanation and instability of the parallel light of instrument and non-uniformity of pixel photoelectric response of linear CCD were discussed. Combined with the measurement principle of the instrument, the calibration method of dynamic threshold setting and the optical glass stick sampling was put forward by theoretical analysis and experimentation. This method can obtain the calibration coefficients of each pixel in different scale particle. The experimental results show that the error of the automatic optical measurement instrument of precipitation can be less than ±4% after calibration. This method is easily realized and has a high feasibility.