Minimizing Matrix Effect by Femtosecond Laser Ablation and Ionization in Elemental Determination.

Bochao Zhang,Miaohong He,Wei Hang,Benli Huang
DOI: https://doi.org/10.1021/ac400072j
IF: 7.4
2013-01-01
Analytical Chemistry
Abstract:Matrix effect is unavoidable in direct solid analysis, which usually is a leading cause of the non-stoichiometric effect in quantitative analysis. In this research, experiments were carried out to study the overall characteristics of atomization and ionization in laser-solid interaction. Both nanosecond (ns) and femtosecond (fs) lasers were applied in a buffer-gas-assisted ionization source coupled with an orthogonal time-of-flight mass spectrometer. Twenty-nine solid standards of ten different matrices, including six metals and four dielectrics, were analyzed. The results indicate that the fs-laser mode offers more stable relative sensitivity coefficients (RSCs) with irradiance higher than 7 x 10(13) W.cm(-2), which could be more reliable in the determination of element composition of solids. The matrix effect is reduced by half when the fs-laser is employed, owing to the fact that the fs-laser ablation and ionization (fs-LAI) incurs an almost heat-free ablation process and creates a dense plasma for the stable ionization.
What problem does this paper attempt to address?