Accuracy of Single-Shot Autocorrelation Measurements of Petawatt Laser Pulses

Xiaoping Ouyang,Jingui Ma,Lin Yang,Shunxing Tang,Chong Liu,Yonghua Peng,Liejia Qian,Baoqiang Zhu,Jianqiang Zhu,Zunqi Lin
DOI: https://doi.org/10.1364/ao.51.003989
IF: 1.9
2012-01-01
Applied Optics
Abstract:At the Shen Guang II (SGII) Petawatt Laser Facility, measurements of large-energy, single-shot laser pulses sometimes feature asymmetric autocorrelation signals, causing uncertainty in the measurement of compressed pulses. This study presents a method for defining and describing the asymmetry of autocorrelation signals. We discuss two sources of asymmetry: the nonuniform distribution of the near field excited by a beam, and the rotation of autocorrelator arms from the cylinder lens. The pulsewidth of an asymmetric autocorrelation signal is shorter than its real width. After updating the autocorrelator, the single-shot autocorrelator for the SGII petawatt laser exhibits a measurement uncertainty of below 12.3%. Recommendations on reducing asymmetry in large-energy, single-shot autocorrelation are discussed.
What problem does this paper attempt to address?