Gas stable isotopic ratio mass spectrometry for measuring δ^13C and δ^18O in background atmospheric CO2

刘立新,周凌晞,夏玲君,张芳,顾帅
DOI: https://doi.org/10.13671/j.hjkxxb.2012.06.015
2012-01-01
Abstract:A high precision gas stable isotopic ratio mass spectrometry system was set up to measure CO2 stable isotopes in background air samples. The system consisted of a MAT253 isotopic ratio mass spectrometry, an efficient airtrap extraction line and a custom designed auto sampler. In combination with periodic target injection, two working standards were injected before and after each of the unknown sample to control and monitor the system performance. The working standards were calibrated regularly to trace back higher level reference and to ensure measurement comparability. From repeated test, the system precision was better than 0.03‰ and 0.0‰ for δ^18 C and δ^18 O, respectively. The system could apply to the δ^18 C and δ^18O analysis in background atmospheric CO2 and essentially meet the WMO/GAW quality target.