Theory and Method for Calibration of High-g Accelerometers with a Narrow Shock Pulse

Jinbiao FAN,Yan WANG,Peng XU,Jing ZU,Zusen Lin
DOI: https://doi.org/10.3969/j.issn.1004-1699.2012.09.013
2012-01-01
Abstract:In order to determine the duration of shock pulse excited a high-g accelerometer, a calibration principle was put forward in this paper. For the accelerometers with different natural frequency,this principle presented the relationship of the maximum excited pulse duration and the accelerometer natural frequency used in the dynamic characteristic calibration of a high-g accelerometer. According to the principle,a narrow pulse shock calibration system was designed and a set of calibration tests was conducted for a high-g piezoelectric accelerometer 8309. It was demonstrated that this principle is effective for amplitude-frequency response characteristic calibration of a high-g accelerometer and various resonance frequencies of a high-g accelerometer were excited using a shock pulse whose duration is determined by the above principle.