Application of Laser Differential Confocal Technique in Back Vertex Power Measurement for Phoropters

Fei Li,Lin Li,Xiang Ding,Wenli Liu
DOI: https://doi.org/10.1117/12.975924
2012-01-01
Abstract:A phoropter is one of the most popular ophthalmic instruments used in optometry and the back vertex power (BVP) is one of the most important parameters to evaluate the refraction characteristics of a phoropter. In this paper, a new laser differential confocal vertex-power measurement method which takes advantage of outstanding focusing ability of laser differential confocal (LDC) system is proposed for measuring the BVP of phoropters. A vertex power measurement system is built up. Experimental results are presented and some influence factor is analyzed. It is demonstrated that the method based on LDC technique has higher measurement precision and stronger environmental anti-interference capability compared to existing methods. Theoretical analysis and experimental results indicate that the measurement error of the method is about 0.02m(-1).
What problem does this paper attempt to address?