Diatom based biosensor for high sensitive fluorescence detection based on a spin-on glass bonding technique

Junfeng Pan,Jun Cai,Deyuan Zhang,Yonggang Jiang,Yu Wang,Mingli Chen,Aobo Li
DOI: https://doi.org/10.1109/ICSENS.2012.6411218
2012-01-01
Abstract:Diatom, with delicate nanoscale porous SiO2 structures, has a promising application in biosensing. We bond diatoms with SiO2 substrate using spin-on glass (SOG) as interlayer to acquire a high performance diatom based fluorescence detecting biosensor. As diatoms can improve proteins density, diatom based biosensors have higher sensitivity compared with traditional protein sensors. This biosensor is fully made of SiO2 which can be modified for biomass attachment. Nanoscale structures of bonded diatoms are better preserved than any related works. The highest bonding strength is 0.6 MPa. Human immunoglobulin G (IgG) is attached on traditional protein sensors and diatom based biosensors as antibody. After hybridized with Cy3-goat anti-human IgG, the diatom based biosensor provides a fluorescence detection signal 2.5 times higher than the traditional one. Given improved fluorescence detecting sensitivity, this new kind of diatom based biosensor can be widely used for various protein detections.