A Calculation Method to Improve the Accuracy of N-Phantom Based Probe Calibration

Li-Ren ZHU,Wen-Jun LI,Hui DING,Guang-Zhi WANG
DOI: https://doi.org/10.3969/j.issn.0258-8021.2012.03.04
2012-01-01
Abstract:In ultrasound guided intervention, in order to track the ultrasonic imaging plane using a 3D spatial tracking device, a transform between ultrasound probe and the attached sensor of the tracking system has to be determined by calibration. The accuracy of calibration has significant impact on the overall accuracy of ultrasound guidance. The widely used N-phantom and traditional least square (LS) calculation method are troubled by lost of coplanar property of registration point set, and may not produce an accurate result. In this paper, a calculation method was presented to tackle this problem, and improve the accuracy of N-phantom based calibration. This calculation method firstly reconstructed the planarity of input point set, and then computed the transform using an LS algorithm with coplanar constraint. Calibration experimental results showed that the new method decreased calibration error to 4.11 mm compared with 5.88 mm of the conventional method.
What problem does this paper attempt to address?