3P-0872 Reproducibility of carotid atherosclerotic lesion type determination using high-resolution magnetic resonance imaging

B. Chu,T.S. Hatsukami,X. Zhao,N.L. Polissar,J.C. Waterton,J. Raichlen,W. Hamar,C. Yuan
DOI: https://doi.org/10.1016/S1567-5688(03)91090-7
2003-01-01
Abstract:We have developed a fast calibration method for computing the position and orientation of 2-D ultrasound (US) images in 3-D space where a position sensor is mounted on the US probe. This calibration is required in the fields of 3-D ultrasound and registration of ultrasound with other imaging modalities. Most of the existing calibration methods require a complex and tedious experimental procedure. Our method is simple and it is based on a custom-built phantom. Thirty N-fiducials (markers in the shape of the letter “N”) embedded in the phantom provide the basis for our calibration procedure. We calibrated a 3.5-MHz sector phased-array probe with a magnetic position sensor, and we studied the accuracy and precision of our method. A typical calibration procedure requires approximately 2 min. We conclude that we can achieve accurate and precise calibration using a single US image, provided that a large number (approximately ten) of N-fiducials are captured within the US image, enabling a representative sampling of the imaging plane. (E-mail: [email protected])