Determination Of Formaldehyde In Glyphosate Technical By Monolithic C-18 Flash Column Separation And Ultraviolet Detection

W Zhang,Wm Li,By Huang,Cp Pan
DOI: https://doi.org/10.3321/j.issn:0253-3820.2005.08.019
2005-01-01
Abstract:An analytical method for determination of formaldehyde impurity in glyphosate technical by offline derivatization-high performance liquid chromatography ( HPLC) was developed. The sample was resolved in acetonitrile-water and derivatized with 2,4-dinitrophenyl hydrazine at pH 2.0 and then extracted under ultrasonic bath. The derivatized product was separated on a Merk Co. monolithic flash column (4. 6 mm x 150 mm) with mobile phase consisting of acetonitrile : water (55:45, V/ V), flow rate of 3 mL/min and detected by ultraviolet detection at 360 nm. Formaldehyde was determined by a reversed phase HPLC monolithic flash column, off-line derivatization with hydrazine reagent and ultraviolet (UV) detection at 360 rim. The method was shown to be simple, rapid and practical: The method was validated from 2.0 to 200 mg/L with satisfied linear relationship. The recovery rates were between 88% to 105%, with less than 5% of relative standard deviation. The detection limit is 0. 5 mg/L. A comparison was undertaken for regular C-18 and monolithic column in separation time and limit of detection ( LOD). With the monolithic column, lower LOD can be obtained, and the analytical time reduced to 1. 5 minutes.