The Effect of Parameter on Determination by HPLC UV_vis Detector

Chu Hongtao,Zhang Qinghe,Li Tong,Zhang Weibing
DOI: https://doi.org/10.3969/j.issn.1003-8892.2005.06.022
2005-01-01
Abstract:The selection of operating parameters for the high performance liquid chromatograph affects the experimental results.In this paper,the effects of parameter settings of the instrument on the analysis results is studied by taking a UV 229 type high precision UV-VIS detector for high performance liquid chromatography as an object.The parameters studied include the width of narrow,response time and detecting wavelength.The results are evaluated based on the performance of detection limit,linear scale and signal/noise ratio.It shows that larger width of narrow deteriorates the linear scale of detector,but improves the detecting sensitivity and reduces detection limit.Increase of response time is advantageous for the control of the noise in some extent,but makes the peak lower and wider especially for the early-eluted peaks.It is very important to select suitable detection wavelength for different samples,which cannot only improve the detection sensitivity,but also enlarges the linear scale of detector.In the establishment of practical analysis method,the instrumental parameters can be adjusted according to the purpose of analysis in order to obtain ideal analysis results.
What problem does this paper attempt to address?