Clinical study of left ventricular ejection fraction measured from gated myocardial perfusion SPECT with ~(99)Tc~m-MIBI by ECToolbox software

LI He-bei,WANG Qian,YUE Ming-gang,WANG Yu,ZHANG Cai-qun,LIANG Tie-jun,NIE Yu-xin,ZHAO Ya-mei
DOI: https://doi.org/10.3969/j.issn.1002-266X.2010.36.006
2010-01-01
Abstract:Objective To evaluate clinical value of ECToolbox software measuring left ventricular ejection fraction(LVEF) by gated myocardial imaging.Methods Rest gated myocardial SPECTS with 99Tcm-MIBI was performed on fifty-eight patients by ECToolbox software.All patients were divided into three groups according to EDV.R0,R1 and R2 were recorded.Equilibrium gated blood-pool imaging was also performed within a week.The correlation and concordance of the two methods measuring LVEF were observed.Results When EDV<70 ml,there were no correlations between the two methods.The LVEF measured by ECToolbox software was significantly higher than that of gated blood-pool imaging.R1 was better.When EDV was between 70 and 100 ml,there was positive correlation between two methods and there was no significant difference of LVEF measured by R1 and gated blood-pool imaging.When EDV≥100 ml,there was strong correlation between the two methods and there was no significant difference of LVEF measured by R0 and gated blood-pool imaging.Conclusions The correlations and concordances of ECToolbox software and gated blood-pool imaging when measuring LVEF were all affected by EDV.R1 should be used when EDV was smaller than 100 ml,otherwise R0 should be used.