Separation and Determination of Chromium(Ⅲ) and Chromium(Ⅵ) by Capillary Zone Electrophoresis

WANG Bai-song,DING Fu-xin,WANG Wu-shang
DOI: https://doi.org/10.3969/j.issn.1004-4957.2006.06.015
2006-01-01
Abstract:Capillary zone electrophoresis(CZE) was applied to the separation and determination of chromium(Ⅵ) and chromium(Ⅲ) with formate as background electrolyte.Cr~(3+) was transformed into ~-by using cyclohexane-1,2-diaminetetraacetic acid(CDTA) as complexing reagent.The relative standard deviations of the determinations of migration times(n=9) for ~-and Cr(Ⅵ) were 0.5% and 0.8%,respectively,and those of peak areas for ~-and Cr(Ⅵ) were 4.6 and 3.0,respectively.The detection limits for signal-to-noise ratio of 2 were 0.08 mg/L for ~-and 0.3 mg/L for Cr(Ⅵ).The linear ranges of peak area vs.concentration were 0.15-10 mg/L with a correlation coefficient of 0.999 7 and 0.625-20 mg/L with a correlation coefficient of 0.998 6 for ~-and Cr(Ⅵ),respectively.The method is simple and feasible.
What problem does this paper attempt to address?