Abstract:An automated two-column ion-exchange system was developed to determine the speciation of chromium in water samples. The system is based on retention of cationic Cr(III) species by Chelex-100 resin and anionic Cr(VI) species by an anion resin. The chromium species retained by the resins are eluted sequentially with stripping reagents and are detected by an on-line flame atomic absorption spectrophotometer. Operation of valves and pumps, data acquisition, and data manipulation are controlled by a microcomputer. The detection limit with a 2 ml sample loop is 2 ng ml−1 for Cr(III) and Cr(VI). The fraction of Cr(III) retained by the Chelex-100 varies with the pH of the solution. In the presence of humic acid, Cr(III) forms negatively charged complexes, some of which are retained by the anion resin.
chemistry, analytical
What problem does this paper attempt to address?