Model-Based Tungsten Anode X-ray Tube Spectra Reconstruction Method

YANG Ying,MOU Xuan-qin,YU Hou-jun,CHEN Xi,ZHANG Yan-bo,TANG Shao-jie
2010-01-01
Abstract:Knowledge of X-ray spectra is essential for X-ray dose control,X-ray image quality evaluation and dual-energy X-ray imaging etc.A model-based tungsten anode X-ray tube spectra reconstruction method is presented in this paper.A physical meaningful seven-parameter X-ray tube spectra model is firstly established.Then model parameters are estimated from attenuation data so that X-ray tube spectra can be reconstructed.In the experiment,a series of aluminum plates and copper plates with different thickness are selected as attenuator for a selected tube voltage.Model parameters are estimated from experimentally measured attenuation data to reconstruct X-ray tube spectra.Experiments show that the method is easy to apply and can reconstruct both bremsstrahlung and characteristic spectra with only a few attenuation data.The difference of attenuation data between initial spectra and reconstructed spectra is lower than 0.3%.The difference between modified handbook spectra and reconstructed spectra is lower than 5%.
What problem does this paper attempt to address?