High Performance Liquid Chromatographic Method For The Determination Of Apigenin And Selected Flavonoids In Human Plasma Using Electrochemical Detection
Zhao Xiu-Juan,Yan Zhao,Zhang Yu-Qiu,Kun Wu
DOI: https://doi.org/10.3321/j.issn:0253-3820.2007.10.028
IF: 1.193
2007-01-01
Chinese Journal of Analytical Chemistry
Abstract:A method was developed for the determination of apigenin and selected flavonoids in human plasma by reversed phase high performance liquid chromatography. Plasma samples were extracted with selective solid phase extraction columns. Separation and quantification were achieved by a Hypersil C-18 column under the optimum conditions: methanol-aqueous phosphoric acid (2/3, VIV, pH 2.25) as mobile phase at a flow rate of 1 mL/min, 40 degrees C and detected with electrochemical detector at 1100 mV. The quantified analysis of internal standard calibration curves was used. The calibration graph was linear in the ragner of 5 - 7000, 3 - 5900, 3 - 7000, 5 - 7000 and 3 - 7400 nmol/L with r > 0.99 for quercetin, luteolin, kaempferol, apigenin, isorhamnetin, respectively. Limits of detection varied from 1.4 to 4.8 -nmol/L. For spiked plasma analysis, the RSD was less than 7.4% and the recoveries were 86.8% - 102.9%. This method is simple, fast, accurate and convenient, and can be applied to the human plasma samples from interventional studies as well as observational studies.