Methodology for gas chromatographic–mass spectral analysis of volatile organic compounds emerging from a low-pressure, flow-through reaction cell

Cheng Yu Ma,Dennis L McCorkle,Weixing Ding,Lal A Pinnaduwage
DOI: https://doi.org/10.1016/S0021-9673(99)00270-8
IF: 4.601
1999-01-01
Journal of Chromatography A
Abstract:A methodology has been developed for the analysis of volatile organic compounds (VOCs) emerging from a plasma discharge cell operating at a constant flow-rate under subatmospheric pressure (0.266 to 2.66 kPa). The analytical system consisted of a gas reservoir for trapping a portion of the VOC–rare gas mixture, a sampling loop for cryogenically concentrating the VOC products, and either gas chromatography–mass spectrometry (GC–MS) or gas chromatography–flame ionization detection (GC–FID). The methodology was evaluated for the analysis of methylene chloride, benzene and tetrachloroethylene, using n-octane as the internal standard. Calibration curves were constructed by plotting the pressure ratios of the gas standard relative to the internal standard versus the corresponding peak area ratios. Over a pressure range of 1.133 to 5.32 kPa, the linearity of the calibration curve for each gas standard was determined with correlation coefficients ranging from 0.96 to 0.98. The relative standard deviation for a minimum of triplicate analyses varied from 1.1 to 18.3% for most VOCs. The calibration curves were used to measure the concentration of premixed VOC–rare gas mixtures as a function of energy input of the plasma reactor.